HKV4000 High Temperature Kinematic Viscosity Bath with Integrated Digital Timing
HKV4000 High Temperature Kinematic Viscosity Bath with Integrated Digital Timing
Description
Connection of the built-in cooling coil to tap water or a recirculating water chiller facilitates temperature control at ambient or below-ambient temperatures. Communications software (RS232, etc.) ramp-to-set, and other enhanced features are available at an additional cost.
Integrated Timing Features -The user can enter the viscosity constant for each viscometer on the front Touch Screen Control Panel, and then get the test result in both efflux time and viscosity units automatically after stopping each timer. Timing function is displayed in 0.01 second resolution and is accurate within 0.01%.
Bath Construction and Safety Features - The bath chamber is a clear borosilicate glass vessel enclosed in a polyester-epoxy finished steel housing. Top working surface has seven 2" (51mm) viscometer ports. Front viewing window assures safe, distortion-free viewing. Microprocessor temperature controller incorporates safety circuitry that interrupts power to the heaters in the event of an overtemperature condition or disconnection of the primary probe. For added safety, an adjustable redundant controller with separate sensor probe interrupts power if an overtemperature situation occurs. An integrated low-liquid sensor prevents operation of the bath if the bath liquid is not filled to the proper level and cuts off power should it fall below during operation. Both over-temperature and low liquid level circuits will latch and prevent further operation of the bath until the fault is removed.
Features
- Microprocessor temperature control between ambient and 232 °C (450 °F)
- Integrated digital timing for easy measurement of sample efflux times
- Permits entry of viscometer constants for automatic calculation and display in viscosity units or seconds
- Dual digital displays show setpoint and actual bath temperature
- Selectable temperature scale - Fahrenheit or Celsius
- Conforms to ASTM D445, D2170 and related specifications
Specifications
Conforms to the specifications of: ASTM D445, D2170, D6074, D6158; IP 71, 319; ISO 3104; DIN 51550; FTM 791-305; NF T 60-100
Temperature Control Range: ambient to 232°C (450°F); sub-ambient to 10°C with external cooling
Display: 0.1 °C / 0.1 °F resolution, calibrate to 0.01 °C / 0.01 °F
Control accuracy and uniformity: Exceeds ASTM requirements throughout the operating range
Timing: Integrated timing function within Touch Screen Control Panel displayed to 0.1 seconds and accurate within 0.01%
Communication: RS232, RS485 port
Viscometer ports: Seven round 2" (51mm) ports
Bath Medium: Water or suitable heat transfer fluid
Bath Capacity:
12": 5.8 gal (22L)
18": 8.9 gal (33.7L)